Fluke FLK-729 300G -0.82 bar to 20.68 bar Pressure Calibrator
- RS Stock No.:
- 144-5554
- Mfr. Part No.:
- FLK-729 300G
- Manufacturer:
- Fluke
Currently unavailable
Sorry, we don't know when this will be back in stock.
- RS Stock No.:
- 144-5554
- Mfr. Part No.:
- FLK-729 300G
- Manufacturer:
- Fluke
Specifications
Technical data sheets
Legislation and Compliance
Product Details
Find similar products by selecting one or more attributes.
Select all | Attribute | Value |
|---|---|---|
| Brand | Fluke | |
| Product Type | Pressure Calibrator | |
| Maximum Pressure Measurement | 20.68bar | |
| Battery Type | Rechargeable | |
| Minimum Pressure Measurement | -0.82bar | |
| Accuracy | ±0.01 % | |
| Resolution | -0.8273 to 20.6843 bar | |
| Pressure Port Type | Differential | |
| Model Number | FLK-729 300G | |
| Data Logging | Yes | |
| Display Type | LCD | |
| Current Loop Measurement | Yes | |
| Leak Testing | Yes | |
| Power Source | Battery | |
| Minimum Operating Temperature | -10°C | |
| Weight | 2.95kg | |
| Maximum Operating Temperature | 50°C | |
| Height | 7cm | |
| Width | 27.94cm | |
| Standards/Approvals | No | |
| Length | 172.7mm | |
| Hazardous Area Certification | NO | |
| Select all | ||
|---|---|---|
Brand Fluke | ||
Product Type Pressure Calibrator | ||
Maximum Pressure Measurement 20.68bar | ||
Battery Type Rechargeable | ||
Minimum Pressure Measurement -0.82bar | ||
Accuracy ±0.01 % | ||
Resolution -0.8273 to 20.6843 bar | ||
Pressure Port Type Differential | ||
Model Number FLK-729 300G | ||
Data Logging Yes | ||
Display Type LCD | ||
Current Loop Measurement Yes | ||
Leak Testing Yes | ||
Power Source Battery | ||
Minimum Operating Temperature -10°C | ||
Weight 2.95kg | ||
Maximum Operating Temperature 50°C | ||
Height 7cm | ||
Width 27.94cm | ||
Standards/Approvals No | ||
Length 172.7mm | ||
Hazardous Area Certification NO | ||
RoHS Status: Exempt
Fluke Pressure Calibrator, ±0.01% Accuracy, 20.68 Bar Maximum Pressure Measurement - FLK-729 300G
This pressure calibrator is a portable, battery-powered instrument designed for pressure generation, control and measurement in process environments. It combines automatic electric pumping with pressure regulation and supports electrical loop functions including mA sourcing, simulation and measurement, enabling field technicians to perform multi-point calibrations and document results efficiently.
Features & Benefits
• Automatic pump to 20.68 bar for set-point attainment and control
• ±0.01% accuracy for precise measurement and calibration results
• 24V loop power for powering and testing transmitters
• Built-in leak testing to verify integrity of test setups
• LCD multi-channel graphical display for clear data visualisation
• Rechargeable battery operation for field portability and autonomy
• ±0.01% accuracy for precise measurement and calibration results
• 24V loop power for powering and testing transmitters
• Built-in leak testing to verify integrity of test setups
• LCD multi-channel graphical display for clear data visualisation
• Rechargeable battery operation for field portability and autonomy
**Applications**
• Suitable for commissioning and calibration of process transmitters
• Ideal for loop current measurement verification on plant skids
• Used for HART transmitter mA trim and light configuration tasks
• Can be used for pressure switch testing with automated sequences
• Used with RTD probes for temperature-related calibration checks
• Ideal for loop current measurement verification on plant skids
• Used for HART transmitter mA trim and light configuration tasks
• Can be used for pressure switch testing with automated sequences
• Used with RTD probes for temperature-related calibration checks
What pressure range can it cover for industrial calibrations?
It can apply and measure from -0.82 to 20.68 bar, accommodating vacuum through high-pressure tasks.
How does it simplify multi-point calibration workflows?
You enter start/end pressures and point count and the device automatically steps, stabilises and logs each test point.
What electrical measurement capabilities does it provide beyond pressure?
It measures, sources and simulates 0–24mA, measures 0–30Vdc and supplies 24V loop power for standalone tests.
How does the instrument handle minor system leaks during a test?
An automatic fine-pressure adjustment feature compensates for small leaks to maintain set-points without manual vernier adjustments.
What environmental operating limits affect pressure control and charging?
Measurement functions operate from -10°C to 50°C while pressure control is rated 0°C to 50°C and the battery will only charge up to 40°C.
Related links
- Fluke FLK-729 150G -0.82 bar to 10.34 bar Pressure Calibrator
- Fluke FLK-729 30G -0.82 bar to 2.06 bar Pressure Calibrator
- Fluke FLK-729 30G FC -0.82 bar to 2.06 bar Pressure Calibrator
- Fluke 717 -850 mbar to 20.68 bar Pressure Calibrator
- Fluke 718 -850 mbar to 20 bar Pressure Calibrator
- Fluke 721EX -0.83 bar to 6.9 bar Pressure Calibrator
- Fluke 719PRO -0.8 bar to 2 bar Pressure Calibrator
- Fluke 719PRO -0.8 bar to 10 bar Pressure Calibrator