RTD, TC Temperature Transmitter, 7.5 → 30 V dc

Technical data sheets
Legislation and Compliance
RoHS Certificate of Compliance
COO (Country of Origin): DK
Product Details

5437 HART 7 Single Input Temperature Transmitter

PR Electronics 5437 single input temperature transmitter offers unrivalled performance in the most demanding of applications to convert a wide array of input types to a 4 – 20 mA output signal. The transmitter has 4 input terminals allowing measurement of two, three or four wire RTDs. Configuring the transmitter with either HART 7 or HART 5 protocols provide flexibility and future proof communication.

Sensor Inputs

The PR 5437 provides unrivalled input versatility. Virtually any thermocouple or RTD plus linear resistance up to 100 kΩ, potentiometer, and bipolar mV signals. Input sensor redundancy function maintains process availability and safety in the event of primary sensor failure. An alert via HART can be used to allow scheduling of maintenance on the faulty sensor.

Installation Characteristics

Wiring - Innovative housing design means the 5437 temperature transmitter can be wired from either the inside or the outside of the terminals.
1/4” centre hole – A large diameter centre hole for easier installation over a wide range of probe mounting diameters.
Loop test pins - Accessible test pins allow measurement of loop current directly while maintaining loop integrity.
Status LED - On-board LED indicates faults according to NAMUR NE44 and NE107.
Jumper pins - Jumpers allow selection of write protection and transmitter burnout options.

Features and Benefits

• LED provides visual notification of sensor, transmitter, and loop status
• HART 7 and HART 5 communication protocols
• Sensor drift detection maintains measuring point reliability
• Flexible input capability, with choice of output modes
• Superior accuracy ensures true, reliable measurements
• Fast response time easily supports safety and cost-critical applications
• Very low temperature coefficient ensuring maximum accuracy across changing ambient conditions
• High galvanic isolation (2.5 kVAC / 55 VAC) providing protection from ground loops, high voltage transients and common mode interference
• Runtime meters provide a valuable history of the process temperature and the ambient temperature around the transmitter

5437D1S

The 5437D can be mounted in zone 0, 1, 2 and zone 20, 21, 22 including M1 / Class I, Division 1, Groups A, B, C, D.

Approvals

SIL Certification
Full third party functional safety assessment to SIL 2 / SIL 3. According to IEC 61508-2010 including firmware assessment to SIL 3. SIL enable/disable function to suit individual applications.

Specifications
Attribute Value
Input Type Linear Resistance, Potentiometer, RTD, Thermocouple, Voltage
Supply Voltage 7.5 → 30 V dc
Diameter 44 mm
Minimum Temperature Input -200°C
Maximum Temperature Input 850°C
Minimum Operating Temperature -40°C
Maximum Operating Temperature +80°C
Series 5437
Depth 20.2mm
Hazardous Area Certification ATEX
1 In Global stock for delivery within 4 - 6 working days
Price Each
MYR 2,472.52
units
Per unit
1 +
MYR2,472.52
Related Products
PR Electronics 6331B intrinsically safe 2-wire programmable temperature ...
Description:
PR Electronics 6331B intrinsically safe 2-wire programmable temperature transmitter offers unrivalled performance in applications making heavy demands on accuracy to convert a wide array of input types to a 4 – 20 mA output signal. The 6331 transmitters provides unrivalled ...
PR Electronics 6333B intrinsically safe 2-wire programmable temperature ...
Description:
PR Electronics 6333B intrinsically safe 2-wire programmable temperature transmitter offers unrivalled performance in applications making heavy demands on accuracy to convert a wide array of input types to a 4 – 20 mA output signal. The 6333 transmitters provides unrivalled ...
4 to 20mA loop powered output and reverse ...
Description:
4 to 20mA loop powered output and reverse polarity protection, configurable without the need for a PC or software. Digital in-head transmitterPush button configuration in less than 10 secs Drift free digital technologyOperating Temperature Range: 0°C to +70°CDimensions: H21mm x ...
Analogue temperature transducer for PT100 sensors and type ...
Description:
Analogue temperature transducer for PT100 sensors and type K thermocouplesScrew terminals, clamping range 0.13 → 1.5 mm2Sensor break and short-circuit detectionEMC according to EN 50081-1 and EN 50082-2.