Phoenix Contact Current Sensor, 4 → 20mA output current

Technical data sheets
Legislation and Compliance
RoHS Certificate of Compliance
COO (Country of Origin): JP
Product Details

Current Transducers - MCR-SL-S

MCR-SL-S series current measuring transducers for measuring sinusoidal and non-sinusoidal alternating currents.

Loop-powered
Can be retrofitted with open-up Rogowski coil
Measuring range selection via slide switch
Operating temperature: -20 ºC to 60 ºC
Pluggable screw connection

Specifications
Attribute Value
Input Current Range 0 → 400 A
Output Current Range 4 → 20mA
Supply Voltage 20 → 30 V dc
Maximum Cable Diameter 18.5mm
Mounting Type DIN Rail
Frequency Bandwidth 30 → 6000 Hz
Accuracy ±1 %
Minimum Operating Temperature -20°C
Maximum Operating Temperature +60°C
Overall Height 85mm
Overall Width 55mm
Overall Depth 70.5mm
Thermal Drift 0.025%/K
Discontinued product
Related Products
The Honeywell CNS series of closed loop Hall-effect ...
Description:
The Honeywell CNS series of closed loop Hall-effect current sensors utilise the null balance or zero magnetic flux feedback method to provide accurate and repeatable measurements. The devices produce an output current which can be converted to a voltage by ...
Measurement of dc currentsLarge apertureEasy panel mountingNo insertion ...
Description:
Measurement of dc currentsLarge apertureEasy panel mountingNo insertion lossesSupply voltage 20/50Vdc - 22/38VacAccuracy ±1%Size (HxWxD) 57.2 x 30 x 89.7mmOperating Temperature -20 → +50°CWeight 100g.
A wide range of current transducers for the ...
Description:
A wide range of current transducers for the electronic measurement of current dc, ac, pulsed... with galvanic separation between the primary and the secondary circuit. Current outputInput and output connections with M5 studsCompatible with AV 100 familyCompact designLow consumption and ...
MCR-SL-CUC series universal current transducer for measuring dc, ...
Description:
MCR-SL-CUC series universal current transducer for measuring dc, ac, and distorted currents. Universal current measurement, no shunt requiredCompact dimensions also enable distributed useVariable mounting on DIN rail and mounting plateSimple connection method due to COMBICON plug-in connection terminal blocks.